Pooja Bakshi, Andrew Korey, William Fowler, Ajay K. Banga. Development and validation of an HPLC-UV method for analysis of methylphenidate hydrochloride and loxapine succinate in an activated carbon disposal system[J]. Journal of Pharmaceutical Analysis, 2018, 8(6): 349-356.
Citation: Pooja Bakshi, Andrew Korey, William Fowler, Ajay K. Banga. Development and validation of an HPLC-UV method for analysis of methylphenidate hydrochloride and loxapine succinate in an activated carbon disposal system[J]. Journal of Pharmaceutical Analysis, 2018, 8(6): 349-356.

Development and validation of an HPLC-UV method for analysis of methylphenidate hydrochloride and loxapine succinate in an activated carbon disposal system

  • Publish Date: Dec. 10, 2018
  • Unused medications have the possibility of being abused, causing serious harm to individuals who were not prescribed the drug. The Food and Drug Administration (FDA) recommends the proper disposal of unused prescribed medications to maintain safety and prevent environmental hazards. However, many of the current disposal techniques do not properly address safety. A drug disposal pouch containing granular activated carbon offers a unique disposal method to deactivate residual or expiredmedication in a convenient, effective, and safe manner. A robust and validated method for methylphenidate hydrochloride and loxapine succinate was developed using high-performance liquid chromatography (HPLC) and the deactivation efficiency of the disposal system was tested. Methylphenidate hydrochloride was analyzed on a C18 analytical column (250mm × 4.60mm, 100?) using acetonitrile-water (0.05% (v/v) trifluoroacetic acid) as the mobile phase at a flow rate of 1.0mL/min with a run time of 15min and retention time of 7.8min. Loxapine succinate was separated on a C8100? (250 mm × 4.6 mm, 5 μm) column maintained at 25 °C using a flow rate of 1.0mL/min. The run time was 10min and the retention time of the drug was around 4.6min.Mobile phase was composed of acetonitrile and water (0.3% triethylamine) at pH 3.0 as 40:60 (v/v). Reference standard solutions (100 μg/mL) for both drugs were prepared by dissolving in mobile phases. These methods provide good linearity (R2 = 0.999) over the range of 5–100 μg/mL for methylphenidate hydrochloride and 0.1–100 μg/mL for loxapine succinate. The assay methods were successfully applied to study the deactivation of these drugs.
  • loading
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索

    Article Metrics

    Article views (100) PDF downloads(4) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return